Detector-based absolute radiometric calibration of lamps and reflectance panels
dc.contributor.advisor | Slater, Philip N. | en_US |
dc.contributor.author | Penny, Iain Reynold, 1966- | |
dc.creator | Penny, Iain Reynold, 1966- | en_US |
dc.date.accessioned | 2013-04-03T13:06:19Z | |
dc.date.available | 2013-04-03T13:06:19Z | |
dc.date.issued | 1991 | en_US |
dc.identifier.uri | http://hdl.handle.net/10150/277922 | |
dc.description.abstract | The absolute calibration of remote sensing systems is based upon radiometric standards. The irradiance standards are currently 1000-watt quartz tungsten-halogen lamps that are traced to National Institute of Standards and Technology standards by a middle party at a significant cost. Lamp instability, short lifetime and calibration inaccuracy present problems. A detector-based absolute calibration technique is described here to replace the lamp calibration in the visible portion of the spectrum (approximately 400 to 700 nm). The system utilizes a quantum efficient QED-200 trap detector in an automated facility to absolutely characterize lamp spectral irradiances or reflectance panel spectral radiances. The measured irradiance of a directly-viewed standard lamp compared favorably (within 0.8%) to the original calibration. Standard-reflectance-panel radiance measurements could not be accurately compared (within 5%) to the illuminating standard lamp irradiance because of excessive stray light in the calibration facility. | |
dc.language.iso | en_US | en_US |
dc.publisher | The University of Arizona. | en_US |
dc.rights | Copyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author. | en_US |
dc.subject | Engineering, Aerospace. | en_US |
dc.subject | Physics, Optics. | en_US |
dc.subject | Remote Sensing. | en_US |
dc.title | Detector-based absolute radiometric calibration of lamps and reflectance panels | en_US |
dc.type | text | en_US |
dc.type | Thesis-Reproduction (electronic) | en_US |
thesis.degree.grantor | University of Arizona | en_US |
thesis.degree.level | masters | en_US |
dc.identifier.proquest | 1345378 | en_US |
thesis.degree.discipline | Graduate College | en_US |
thesis.degree.discipline | Optical Sciences | en_US |
thesis.degree.name | M.S. | en_US |
dc.identifier.bibrecord | .b27003267 | en_US |
refterms.dateFOA | 2018-06-16T19:27:21Z | |
html.description.abstract | The absolute calibration of remote sensing systems is based upon radiometric standards. The irradiance standards are currently 1000-watt quartz tungsten-halogen lamps that are traced to National Institute of Standards and Technology standards by a middle party at a significant cost. Lamp instability, short lifetime and calibration inaccuracy present problems. A detector-based absolute calibration technique is described here to replace the lamp calibration in the visible portion of the spectrum (approximately 400 to 700 nm). The system utilizes a quantum efficient QED-200 trap detector in an automated facility to absolutely characterize lamp spectral irradiances or reflectance panel spectral radiances. The measured irradiance of a directly-viewed standard lamp compared favorably (within 0.8%) to the original calibration. Standard-reflectance-panel radiance measurements could not be accurately compared (within 5%) to the illuminating standard lamp irradiance because of excessive stray light in the calibration facility. |