    Laboratory Evaluation of Eleven Image Intensifiers

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Thumbnail
    Name:
    azu_QC_351_A7_no81_w.pdf
    Size:
    10.37Mb
    Format:
    PDF
    Download
    Author
    Cromwell, Richard H.
    Dyvig, Ronald R.
    Issue Date
    1973-08
    Keywords
    Optics.
    Image intensifiers.
    
    Metadata
    Show full item record
    Publisher
    Optical Sciences Center, University of Arizona (Tucson, Arizona)
    Rights
    Copyright © Arizona Board of Regents
    Collection Information
    This title from the Optical Sciences Technical Reports collection is made available by the College of Optical Sciences and the University Libraries, The University of Arizona. If you have questions about titles in this collection, please contact repository@u.library.arizona.edu.
    Abstract
    A number of laboratory tests have been carried out at the image tube laboratory of the Optical Sciences Center on 11 selected image intensifiers in order to compare and evaluate their performance. The electrostatically focused tubes tested are as follows: ITT F-4708 (1 stage), two samples of a Varo 8605 (1 stage), Varo 8605DC (1 stage), ITT F -4724 (3 stages), RCA 8606 (3 stages), and Varo 8606 (3 stages). The magnetically focused tubes are EEV P829D (5 dynodes), ITT F-4089 (1 stage), RCA C33011 (2 stages), and RCA C70021AEP2 (3 stages). The tests included measurements of limiting resolution (of the intensifiers directly and of photographs obtained with the intensifiers), square -wave response, geometrical distortion, shear characteristics in fiber-optic faceplates, photographic speed gain, uniformity of response, light- induced background, dark emission, and detective quantum efficiency (DQE) of intensifier -photographic emulsion combinations. The most significant result of the tests is that a comparison among tubes with a similar number of stages (or gain) showed that the magnetically focused intensifiers were generally superior to the electrostatically focused intensifiers in resolving power, geometrical distortion, and uniformity of response. However, the electrostatically focused tubes exhibited only about one fifth the light- induced background of the magnetic tubes. Also, a mean relationship was found between the limiting resolution of the intensifier output image and the limiting resolution that is recorded on a photograph of the intensifier. Other particularly notable results include (1) the causes and characteristics of several different types of response nonuniformities, (2) the identification of sources of light -induced background, (3) the photographic speed gain required of an intensifier to obtain the highest peak DQE possible and also that required to make the system behave as a "single- photon event" detector, (4) the identification of some especially undesirable characteristics of a potassium chloride transmission secondary emission (TSE) dynode intensifier (EEV P829D), and (5) the performance of three relay lenses for intensifiers.
    Description
    QC 351 A7 no. 81
    Series/Report no.
    Optical Sciences Technical Report 81
    Collections
    Optical Sciences Technical Reports

    entitlement

     
    The University of Arizona Libraries | 1510 E. University Blvd. | Tucson, AZ 85721-0055
    Tel 520-621-6442 | repository@u.library.arizona.edu
    DSpace software copyright © 2002-2017  DuraSpace
    Quick Guide | Contact Us | Send Feedback
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.