• Development of Graphitization of μg-Sized Samples at Lund University

      Genberg, J.; Stenström, K.; Elfman, M.; Olsson, M. (Department of Geosciences, The University of Arizona, 2010-01-01)
      To be able to successfully measure radiocarbon with accelerator mass spectrometry (AMS) in atmospheric aerosol samples, graphitization of small sample sizes ( 50 g carbon) must provide reproducible results. At Lund University, a graphitization line optimized for small samples has been constructed. Attention has been given to minimize the reduction reactor volume and each reactor is equipped with a very small pressure transducer that enables constant monitoring of the reaction. Samples as small as 25 g of carbon have been successfully analyzed, and the mass detection limit of the system has probably not been reached.