Citation
Schoch, H., Bruns, M., Münnich, K. O., & Münnich, M. (1980). A multi-counter system for high precision carbon-14 measurements. Radiocarbon, 22(2), 442-447.Publisher
American Journal of ScienceJournal
RadiocarbonAdditional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
A new 14C detector system containing nine, independently working, CO2 proportional counters is described. The system is designed for a sufficient measuring capacity at a precision level better than sigma = +/- 2 per mil, which requires a counting time of about one week per sample. The size of the installation requires a simple and economic design of counters and electronics. A single anticoincidence shield for all counters consists of five newly developed flat counters. The modern counting rate (52 cpm) is sensitively checked by running Heidelberg sodium carbonate standard samples wth a counting rate of about 10 times modern. A microcomputer (DEC PDP-11/03) is used for data acquisition. Recent developments in laboratory techniques (preparation and gaschromatographic purification of samples) are also reported.Type
Articletext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200009747