Liquid Scintillation Counter Characterization, Optimization and Benzene Purity Correction
Author
McCormac, F. G.Issue Date
1992-01-01Keywords
optimizationmonitoring
hydrogen
tritium
aromatic hydrocarbons
benzene
hydrocarbons
organic compounds
methods
C 14
carbon
dates
isotopes
radioactive isotopes
absolute age
geochemistry
Metadata
Show full item recordCitation
McCormac, F. G. (1992). Liquid scintillation counter characterization, optimization and benzene purity correction. Radiocarbon, 34(1), 37-45.Journal
RadiocarbonAdditional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
In liquid scintillation counting (LSC), small variations in benzene purity can cause 14C pulse-height spectra to move with respect to the counting window. Thus, one must carefully monitor the purity of each benzene sample and apply corrections for spectral shifts. I describe here the techniques used at Queen's University Belfast for deriving correction factors for observed small variations in benzene purity. I also describe the methods used at our laboratory to fine-tune our Quantulus LS counters for high-precision dating. The tuning of the instruments minimizes the effect of fluctuations in gain that may occur during the long counting periods required for high-precision dating. Any remaining influences on efficiency owing to gain changes are corrected for, along with the purity correction, by continuous monitoring of the spectrum produced by the external source.Type