Performance of the Packard 2000 CA/LL and 2250 CA/XL Liquid Scintillation Counters for 14C Dating
Issue Date
1989-01-01Keywords
instrumentsliquid scintillation methods
methods
C 14
carbon
isotopes
radioactive isotopes
absolute age
Metadata
Show full item recordCitation
Cook, G. T., Harkness, D. D., & Anderson, R. (1989). Performance of the Packard 2000 CA/LL and 2250 CA/XL liquid scintillation counters for 14C dating. Radiocarbon, 31(3), 352-358.Journal
RadiocarbonDescription
From the 13th International Radiocarbon Conference held in Dubrovnik, Yugoslavia, June 20-25, 1988.Additional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
Results are presented on the comparative characteristics of the Packard 2000 CA/LL liquid scintillation counter with and without the low-level option on line. An initial performance assessment using 14C labeled benzene with butyl-PBD as the scintillant revealed that although the background count rate decreased by approximately a factor of 3 using the lowlevel option, a substantial decrease in efficiency was also observed. However, subsequent investigations have indicated that, by careful manipulation of both scintillant composition and concentration, this loss in efficiency can largely be overcome with little or no concurrent increase in background count rate. The introduction of an active vial sample holder and a new light-guide system proved to be significant advances on the standard 2000 CA/LL.Type