Change of Calibration Parameters in Frequently Used Vials in Benzene Scintillation Counting
Citation
Haas, H. (1995). Change of calibration parameters in frequently used vials in benzene scintillation counting. Radiocarbon, 37(2), 737-742.Journal
RadiocarbonAdditional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
Scintillation counting vials used for counting synthesized benzene accumulate an inside coating which is insoluble in benzene. This coating reduces the photon transfer from the vial and lowers its observed background. Age calculations that include calibration data collected with other vials may be inaccurate if this effect is not considered. Frequent cleaning of the counting vials with strong acid solutions keeps the coating from accumulating. A sample counting vial cleaned only with photospectrometric grade benzene between 67 different sample benzene fillings made in a 20-month period showed, during counting, an efficiency lowered by 2.8% and a background lowered by 8.7%. Age calculations for a 10 ka old sample made with calibration data from "matched" vials would yield a result too old by 515 yr.Type
Articletext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200031271