Low-Level Liquid Scintillation Counter Array with Computerized Data Acquisition and Age Calculation Capabilities for 14C Dating
Citation
Noakes, J. E., Spaulding, J. D., & Valenta, R. J. (1995). Low-level liquid scintillation counter array with computerized data acquisition and age calculation capabilities for 14C dating. Radiocarbon, 37(2), 773-779.Journal
RadiocarbonAdditional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
We describe a two-phase study directed toward background reduction of a manual liquid scintillation counter and the interfacing of electronics for counting to a computer data acquisition system. Counter background reduction is achieved with afterpulse electronics, a high-performance cocktail, an auxiliary detector/guard and a special sample vial holder. The data acquisition system is comprised of an electronic signal processor and sorter for operating up to eight counters simultaneously and interfacing to a computer with software for data storage, acquisitions and age dating calculations. We discuss low-background counter modifications, electronic signal processing and computer software for 14C age dating.Type
Articletext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200031325