Citation
Bronk Ramsey, C., & Hedges, R. E. M. (1995). Radiocarbon with gas chromatography. Radiocarbon, 37(2), 711-716.Journal
RadiocarbonAdditional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
In 14C tracer studies, and when looking for modern contamination in archaeological samples, it is often ncessary to measure the 14C concentration of individual chemical fractions. Gas chromatography (GC) is one method that is frequently used for separation of chemical fractions. The gas ion source at the Oxford Radiocarbon Accelerator Unit for accelerator mass spectrometry (AMS) provides the opportunity to measure fractions from a GC instrument directly. Although the first investigations are likely to be 14C tracer studies, such a GC-AMS system could find much wider application. We present results from a pilot study of the peak sensitivity, baseline stability and crosstalk of the accelerator system used in this way. We also discuss the practical considerations in developing a GC-AMS instrument for routine use.Type
Articletext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53