Issue Date
1995-01-01Keywords
Groningen Netherlandsaccelerator mass spectra
instruments
mass spectra
spectra
techniques
Netherlands
Europe
Western Europe
methods
C 14
carbon
isotopes
radioactive isotopes
absolute age
Metadata
Show full item recordCitation
Gottdang, A., Mous, D. J. W., & van der Plicht, J. (1995). The HVEE 14C system at Groningen. Radiocarbon, 37(2), 649-656.Journal
RadiocarbonAdditional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
Since May 1994, a new-generation accelerator mass spectrometer (AMS) has been fully operational at the Centre for Isotope Research in Groningen, The Netherlands. The fully automated and high-throughput accelerator mass spectrometry (AMS) system, manufactured by High Voltage Engineering Europa (HVEE) is dedicated to radiocarbon analysis. The HVEE 4130 14C AMS is able to analyze up to 3000 samples per year. The system is characterized by simultaneous transport of all three isotopes (12C, 13C,14C) and 14C analysis with a precision below 0.5 pMC and a daily stability below 0.5 pMC. We present here a system description together with stability and performance measurements.Type
Articletext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200031167