Citation
Ruff, M., Wacker, L., Gäggeler, H. W., Suter, M., Synal, H.-A., & Szidat, S. (2007). A gas ion source for radiocarbon measurements at 200 kV. Radiocarbon, 49(2), 307-314.Journal
RadiocarbonDescription
From the 19th International Radiocarbon Conference held in Keble College, Oxford, England, April 3-7, 2006.Additional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
The novel tabletop miniaturized radiocarbon dating system (MICADAS) at ETH Zurich features a hybrid Cs sputter negative ion source for the measurement of solid graphite and gaseous CO2 samples. The source produces stable currents of up to 6 A C out of gaseous samples with an efficiency of 36%. A gas feeding system has been set up that enables constant dosing of CO2 into the Cs sputter ion source and ensures stable measuring conditions. The system is based on a syringe in which CO2 gas is mixed with He and then pressed continuously into the ion source at a constant flow rate. Minimized volumes allow feeding samples of 330 g carbon quantitatively into the ion source. In order to test the performance of the system, several standards and blanks have successfully been measured. The ratios of 14C/12C could be repeated within statistical errors to better than 1.0% and the 13C/12C ratios to better than 0.2%. The blank was 1 pMC.Type
Proceedingstext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200042235