Background Components of a Liquid Scintillation Counter in the 14C Window
Citation
Jonsson, G., & Theodórsson, P. (2007). Background components of a liquid scintillation counter in the 14C window. Radiocarbon, 49(2), 315-323.Journal
RadiocarbonDescription
From the 19th International Radiocarbon Conference held in Keble College, Oxford, England, April 3-7, 2006.Additional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
We present a broad and detailed study of the background components of a liquid scintillation (LS) detector, using a simple laboratory-built system, ICELS. It was specifically designed for radiocarbon dating and is compact and easily transportable (total weight 35 kg). Its flexible LS detector unit has a dome-shaped vial with 3 mL benzene to which 45 mg butyl-PBD is added. The vial sits on the top of a vertical 28-mm-diameter phototube. The gamma radiation, to which the benzene is exposed under varying conditions, was measured by replacing the vial with a 38-mm-diameter NaI crystal. The pulse-height spectra of the 14C LS background and the NaI gamma background were measured in a surface laboratory and in a deep underground counting room with: 1) a lead shield of varying thickness; 2) lead of normal and low 210Pb concentration; 3) phototubes of 2 different types; and 4) varying benzene volume. The beta emission from the face of the tubes was measured with a low-level Geiger counter.Type
Proceedingstext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200042247