Gas Counting System for 14C Dating of Small Samples in the Kraków Laboratory
Issue Date
1998-01-01Keywords
gaseous phasecopper
Poland
background level
laboratory studies
precision
size
errors
liquid scintillation methods
Central Europe
metals
experimental studies
Europe
charcoal
methods
C 14
carbon
isotopes
radioactive isotopes
carbon dioxide
absolute age
Metadata
Show full item recordCitation
Gorczyca, Z., Jeleń, K., & Kuc, T. (1998). Gas counting system for 14C dating of small samples in the Kraków laboratory. Radiocarbon, 40(1), 129-135.Journal
RadiocarbonDescription
From the 16th International Radiocarbon Conference held in Gronigen, Netherlands, June 16-20, 1997.Additional Links
http://radiocarbon.webhost.uits.arizona.edu/Abstract
The application of traditional gas or liquid scintillation counting (LSC) is necessary for assessing radionuclide activity in countries without operating accelerator mass spectrometry (AMS) facilities. A simple and relatively inexpensive system of mini gas counters for measurement of radiocarbon in archaeological and environmental samples has been set up recently in the Kraków laboratory (Department of Environmental Physics, University of Mining and Metallurgy). The system is composed of a gas purification and counter filling line, three identical 15-mL copper/quartz counters, active and passive shielding, and an electronic unit with data acquisition. One counter measures 22 mg of carbon as CO2 with efficiency >95% at a background reduced to 0.044 cpm by a NaJ(T1) guard counter and lead shield. The detection limit (1 sigma) for a two-week measurement of 48 mL of CO2 is 0.52 pMC. The corresponding counting error of a 100 pMC environmental sample is 1.3 pMC for 22 mgC (one counter) and 0.75 pMC for 66 mgC (three counters filled with the sample).Type
Proceedingstext
Language
enISSN
0033-8222ae974a485f413a2113503eed53cd6c53
10.1017/S0033822200017963